700 NDIR Analyzer
6-20-2017
145
Serial Interface
The serial interface enables remote control of the analyzer by a master computer. It is
implemented as an RS232 V24 interface and meets all requirements of the AK protocol.
A 9-pin male connector at the back of the unit is used to connect a master computer with the
following pin assignment:
Figure 0-1 Serial Interface
Pin 3 = Txd (transmit)
Pin 2 = Rxd (receive)
Pin 5 = Gnd (ground)
Interface Specifications
Baud Rate:
9600, 4800, 2400, 1200, 600, 300 baud
Data Bits:
7 or 8
Stop Bit:
1 or 2
Don’t Care:
1 byte, adjustable (e.g. 32)
Parity:
Even, odd, none
XON/XOFF:
Active or not active
Hand Shake:
No
Ethernet RJ47
If connecting directly to a computer (without using a hub or switch), a crossover cable is
required.