700 NDIR Analyzer
6-20-2017
45
If the calibration was unsuccessful, the screen will say Outside Deviation Limits.
Example of an unsuccessful calibration.
If the zero calibration is unsuccessful, check the following:
1.
Make sure the correct gas was introduced into the analyzer.
2.
Verify the
Diagnostic Values
while flowing gas.
3.
Check Zero Gas Deviations under
Manual Deviations
:
→
→
→
→
→
4.
Check
Maximum Calibration Error
under Calibration Setup:
→
→
→
→
→
After a successful Manual Zero Calibration, press the
button to return to the Manual
Calibration menu. The analyzer is now ready for a span calibration or check.