700 NDIR Analyzer
6-20-2017
154
A CO2: Query CO2 Correction Parameter
Command
Response
Description
_ACO2_Km
_ACO2_s_Ext1_z.z_y.y_x.x_w.w
Gas correction for Channel m
Ext1: External 1 voltage input
z.z: Offset ( the voltage of analog input with no
correction gas present)
y.y: Min analog input (if analog input is below this
value no gas correction will be done)
x.x: 1
st
order coefficient
w.w: 2
nd
order coeficient
ADAL: Diagnostic Alarm Limits
Command
Response
Description
_ADAL_K0
_ADAL_s_a1.min_a1.max_....
_a16.min_a16max
All alarm limits.
1: Channel 1 Flow
2: Channel 2 Flow
3: Channel 3 Flow
4: External analog 1
5: External analog 2
6: Barometer
7: Analyzer temperature
8: Channel 1 Concentration 1 / Concentration 2
9: Channel 2 Concentration 1 / Concentration 2
10: Channel 3 Concentration 1 / Concentration 2
11: Channel 1 detector temperature
12: Channel 2 detector temperature
13: Channel 3 detector temperature
14: Channel 1 sample EPC voltage %
15: Channel 2 sample EPC voltage %
16: Channel 3 sample EPC voltage %
_ ADAL_K0_x
_ADAL_s_x.min_x.max
Alarm limits of x (x = 1 to 16)