700 NDIR Analyzer
6-20-2017
175
MBAP Header Description
The MBAP Header contains the following fields:
Field
Length
Description
Transaction Identifier
2 Bytes
Identification of a Modbus request / Response transaction
Protocol Identifier
2 Bytes
0 = Modbus
Length
2 Bytes
Number of following bytes
Unit Identifier
1 Byte
Identification of a remote slave connected on a serial line or
on other buses
The header is seven bytes long:
Transaction Identifier
– Used for transaction pairing, the Modbus server copies in the
response the transaction identifier of the request.
Protocol Identifier
– Used for intra-system multiplexing. The Modbus protocol is
identified by the value 0.
Length
- The Length field is a byte count of the following fields, including the Unit
Identifier and Data fields.
Unit Identifier
– This field is used for intra-system routing. It is typically used to
communicate to a or a Modbus serial line slave through a gateway between an
Ethernet TCP-IP network and a Modbus serial line. This field is set by the Modbus client
in the request and must be returned with the same value in the response by the server.
All Modbus/TCP ADU are sent via TCP to Registered Port 502.
The different fields are encoded in Big-endian.
The 700 NDIR/O
2
Analyzer uses only the
Length bytes from the MBAP section.
Содержание 700 NDIR
Страница 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Страница 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Страница 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Страница 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Страница 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...