700 NDIR Analyzer
6-20-2017
111
Output Assignments
→
→
→
→
→
The Output Assignments screen allows the operator to assign any of the 15
programmable digital outputs to a specific alarm or status.
From the Programmable
Digitals menu press
to access the Output Assignments screen.
Use the left and right arrow buttons
to highlight the field you intend to change.
Press
to open the field and use the up or down arrow button
to change
the signal. Press
again to close the field. Press
to save your changes.
NOTES:
1.
For information on digital output connections see
Analog and Digital Interface
.
2.
Alarms go open when present and statuses go closed when active.
3.
See the following tables for a list of Alarms and Statuses: