en_BA_OxE7
_PosCon3D.docx
36/64
Baumer Electric AG
17.07.2014 14:21/tof V1.01 ANW_81125407
Frauenfeld, Switzerland
5.8.3
TEACH REF
Conditions during TEACH REF
The four following conditions must be met during the TEACH REF teach-in process. If one of the symbols
listed below appears on the display, it lights up red. The teach-in process cannot be started until all errors
have been eliminated.
Symbol
Error description
Error correction
Distance between sensor and reference surface
not correct. The distance must be between 180
and 250 mm
Correct distance between sensor and
reference surface
The inclination angle of the sensor to the
reference surface is too large. Maximum
inclination angle ±30°
Correct inclination of the sensor
The reference surface is too uneven. The
unevenness must not exceed ±0.5 mm
Use an auxiliary plate during teach-in
process
The length of the reference surface is too small. It
must have a minimum length of 50 mm
Remove objects from the measuring
field or use an auxiliary plate during the
teach-in process
Start the TEACH REF teach-in process by pressing SET for 2 seconds.
To ensure correct teaching-in of the reference surface, THICKNESS REF must always be completed after the
angle teach-in process. Only in this manner can the effective reference surface be determined with due regard
to the thickness of the auxiliary plate.