8
en
Connecting the instrument
Only instruments of the CAIMAN Seal & Cut product family can be used in
conjunction with the Lektrafuse HF generator GN200.
The instrument connector has an arrow marking and the connection
socket
8
has a dot marking. To connect the plug to the connection socket
correctly, these markings must be aligned.
Fig. 8
►
Connect the instrument with the connection socket
8
.
The green signal ring
7
around the instrument connector lights up. The
message "Ready to Seal" appears in the display
5
.
Fig. 9
Selecting the operating mode
The Lektrafuse HF generator has two operating modes for the sealing pro-
cess. The selected operating mode will be retained until the HF generator
is switched off. The operating mode can be changed during the procedure.
►
Changing the operating mode: Press operating mode selection button
(MODE)
2
.
■
Standard mode: Pre-set operating mode
■
Plus mode: Operating mode with increased energy output
The set operating mode is independent from the connected instrument.
Changing or reconnecting the instrument does not change the set operat-
ing mode.
The active operating mode is indicated in the display as follows:
■
Border
■
Bold font
Fig. 10
Fig. 11
Note
In case of a regrasp error, the active operating mode is not displayed. The
active operating mode can be displayed by pressing the operating mode
selection button (MODE)
2
.
Note
The Plus mode can be distinguished from Standard mode by the higher fre-
quency of the beep during HF discharge.
Decommissioning
►
Switch off the device with the power OFF/ON switch
12
.
All the poles of the device are disconnected from the power supply.
►
Withdraw mains cable.
The device is completely disconnected from the power supply.
4.2
Function checks
Prior to each use, carry out a functional test of the device:
►
Check that the accessories have no visible damage.
►
Prepare and set up the unit, see Preparation and setup.
►
Check the functionality of the following elements, one after the other,
in the following sequence:
– Switch on the power OFF/ON switch
12
; the power on signal lamp
9
lights up
– Automatic self-test after each time the device is switched on: brief
acoustic signal, all display elements light up briefly
– Operating mode selection, see Selecting the operating mode
– Foot control, see Connecting the foot control
– Instrument, see Connecting the instrument
– Activation with instrument button and foot switch
– Switch off the power OFF/ON switch
12
4.3
Safe operation
WARNING
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
the product.
WARNING
Risk of injury when using the product beyond the
field of view!
►
Apply the product only under visual control.
Содержание Aesculap Lektrafuse GN200
Страница 2: ......
Страница 3: ...2 3 1 9 4 6 7 8 5 10 11 12 13 14 15...
Страница 132: ...130 ru Aesculap Lektrafuse 1 2 MODE 3 4 5 6 7 8 Caiman Lektrafuse 9 10 11 2 12 13 14 15 2002 96 EG WEEE CF 93 42 LOT QTY...
Страница 135: ...133 ru Lektrafuse Lektrafuse 12 Lektrafuse 2 2 1 Lektrafuse 2 2 Caiman Lektrafuse GN200 TA022414 TA022130 1 1 5 5 4 0 8...
Страница 141: ...139 ru VAH FDA CE pH 8 www a k i org Rote Brosch re 5 3 6 5 4 5 5...
Страница 142: ...140 ru 5 6 I 5 7 Lektrafuse 5 8 6 Lektrafuse 6 1 B Braun Aesculap T C F t I 1 Meliseptol Rapid 50 1 10 C Lektrafuse 1...
Страница 216: ...214 el 1 1 IEC VDE KIN YNO POEI O OIH H PO OXH PO OXH KIN YNO KIN YNO ON OFF KIN YNO POEI O OIH H POEI O OIH H...
Страница 222: ...220 el 5 4 5 5 KIN YNO PO OXH PO OXH PO OXH...
Страница 223: ...221 el 5 6 I 1 5 7 5 8 6 Lektrafuse GN200 6 1 B Braun Aesculap T C F t min I 1 Meliseptol HBV 50 1 PO OXH 10 C 1...