![Atlas BCX Series Скачать руководство пользователя страница 92](http://html.mh-extra.com/html/atlas/bcx-series/bcx-series_operating-manual_3001603092.webp)
Page 92
BCX Series Operating Manual
The conditions in orange and rose can be selected in conjunction with other conditions
after one condition is selected in the first set.
SPRAY PUMP
Uses a magnetic drive (or similar) pump to pull solution
from the reservoir and send it into the spray bar header for
direct spray (impingement) of the test specimens.
TIME IN SEC.
C
hanges the timer for a specific step from minutes to
seconds.
IMMERSION HOLD SOAK
Holds the cabinet at a specific step in the cycle even if the
step time has elapsed until the programmed immersion
reservoir temperature setpoint has been attained.
USER OUTPUT
Activates an output on the control I/O during that step. The
user can connect a device to that output and have the
device activated during any specified step.
CHAM HOLD SOAK
Holds the cabinet at a specific step in the cycle even if the
step time has elapsed until the programmed cabinet
exposure zone temperature setpoint has been attained.
CHAM RAMP SOAK
Meters power to the cabinet heaters so that the exposure
zone temperature setpoint is achieved at the end of the step
time.
NOTE: Some displayed features are optional and not on each cabinet. They can be used only if the
options were purchased with the cabinet.