BCX Series Operating Manual
H. DIFFUSER
What seems to be a “false floor” of the cabinet is actually an Air Diffuser; its function
being to distribute air flow within the Exposure Zone. Samples or undue weight
shouldn’t be placed upon it.
I. SOLUTION TRANSFER PUMP
This cabinet is equipped with a 35-gallon Solution Reservoir. This is not a mixing tank.
The salt solution should be mixed in another container and transferred into this reservoir.
Due to the location, it is not practical to “pour” the mixed salt water into this reservoir.
To get the pre-mixed salt solution into the reservoir, a couple of simple steps must be
taken.
1. Mix the salt solution in a
separate container.
2. Get this pre-mixed solution
close 1-1.5 m / 3-5’ to the
BCX’s 35-gallon reservoir.
3. Put the tubing from the transfer
pump’s inlet into the container
holding the pre-mixed solution.
4. Change ball valve #3 (Salt
Solution from solution
reservoir) from its normally
OPEN position to a CLOSED
position
5. Change ball valve #4 (air to the
Solution transfer pump) from its
normally CLOSED position to
an OPEN or partially OPEN
position. The more you OPEN
it, the faster the pump will move
solution into this 35-gallon
reservoir.
6. When all of the solution is
moved, put these two valves
into their original positions, coil
up the tubing and the transfer is
complete.
TRANSFER PUMP