Page 112
BCX Series Operating Manual
G. UNEVEN COLLECTION RATES
POSSIBLE
CAUSES CORRECTIVE
ACTION
Worn or clogged atomizer nozzle
Clean and/or replace nozzle
Air pressure setting not accurate
Check air pressure setting
Out-of-calibration pressure gauge
Check gauge calibration
Improper placement of collector funnels
Probably located too high
Improper positioning of adjustable cone
Centrally locate cone height
Throttle holes of Dispersion Tower opened
Close throttle holes
Bubble Tower temperature improper
Check & correct bubble tower temp
Funnel collecting condensation from test specimen Check & correct funnel placement
Empty fog nozzle feed reservoir
Check floats, solenoids & tubing
H. LOW RELATIVE HUMIDITY
POSSIBLE
CAUSES CORRECTIVE
ACTION
Empty fog nozzle feed reservoir
Check floats, solenoids & tubing
RTDs are out of calibration
Check and calibrate RTDs
Controller is out of calibration
Check and calibrate controller
Wet bulb wick & reservoir empty
Clean and fill wick & reservoir
Air pressure setting not accurate
Check air pressure setting
Out-of-calibration pressure gauge
Check gauge calibration
Throttle holes of Dispersion Tower closed
Open throttle holes
Bubble Tower temperature improper
Check & correct bubble tower temp
I. NO HEAT IN BUBBLE TOWER
POSSIBLE
CAUSES CORRECTIVE
ACTION
Activation of the Low Water Cut-off feature
Reset switch on the B.T. heater
Bubble Tower RTD is out of calibration
Check and calibrate RTD
Controller is out of calibration
Check and calibrate controller
Bubble Tower setpoint too low
Check & correct setpoint
Bubble Tower is not full of water
Check & correct water level
Blown fuse for the Bubble Tower heater
Check and replace fuses if needed
Bubble Tower heater is damaged
Check and replace heater if needed