BCX Series Operating Manual
Shown below on the “CURRENT TEST DATA” page is a list of the inputs that can be
plotted.
Since space necessitated an abbreviation of the labels for these inputs, the list below
defines the signal.
Some future inputs planned are “Bubble Tower’s pressure – IN”, “Bubble Tower Air
Flow,” and several spare temperature measurements.
As described on the earlier page, next to the description of the analog input the actual
value for this signal is shown.
Any item listed as a SP (setpoint) is not actually an analog input. This is a value set
by the user on the programming page. These values are listed here to allow the user
the ability of comparing setpoint against actual values in a graphical form.
NOTE: Some described features are optional and not on each cabinet. They will only appear if the
options were purchased with the cabinet.
Abbreviation Definition
Cabinet
The exposure zone’s current actual
temperature in degrees Celsius.
Cabinet SP
The exposure zone’s setpoint in degrees
Celsius for the current step.
Bub Twr
The bubble tower’s current actual
temperature in degrees Celsius.
Bub Twr SP
The bubble tower’s setpoint in degrees
Celsius for the current step.
Lab
The laboratories current actual
temperature in degrees Celsius.
RH
The exposure zone’s current actual
percent relative humidity.
RH SP
The exposure zone’s setpoint for relative
humidity in percentage for the current
step.
BT Pres
The bubble tower’s current actual
pressure in P.S.I. (pounds per square
inch) as it exits the bubble tower (OUT).
Imm Temp
The immersion tank’s current actual
temperature in degrees Celsius.
Imm Temp SP
The immersion tank’s setpoint in degrees
Celsius for the current step.