BCX Series Operating Manual Page 111
D. NO FOG
POSSIBLE
CAUSES CORRECTIVE
ACTION
Clogged atomizer nozzle
Clean or replace nozzle
Clogged solution filter cartridge
Replace filter
No solution in the internal reservoir
Fill reservoir
Inadequate air flow to the nozzle
Check air flow to nozzle
Bubble Tower Air Solenoid Failure
Check/replace solenoid
Customer supplied Compressed Air failure
Check for proper air supply
E. LOW COLLECTION RATES
POSSIBLE
CAUSES CORRECTIVE
ACTION
Clogged atomizer nozzle
Clean or replace nozzle
Clogged solution filter cartridge
Replace filter
Air pressure set too low
Check air pressure setting
Out-of-calibration pressure gauge
Check gauge calibration
Improper placement of collector funnels
Probably located too high
Improper positioning of adjustable cone
Centrally locate cone height
Throttle holes of Dispersion Tower closed
Open throttle holes
Inadequate air flow to the nozzle
Check air flow to nozzle
Customer supplied compressed air failure
Check for proper air supply
Empty nozzle feed reservoir
Check floats, solenoids & tubing
F. HIGH COLLECTION RATES
POSSIBLE
CAUSES CORRECTIVE
ACTION
Worn
atomizer
nozzle Replace
nozzle
Air pressure set too high
Check air pressure setting
Out-of-calibration pressure gauge
Check gauge calibration
Improper placement of collector funnels
Probably located too high
Improper positioning of adjustable cone
Centrally locate cone height
Throttle holes of Dispersion Tower opened
Close throttle holes
Bubble Tower temperature improper
Check & correct bubble tower temp