BCX Series Operating Manual Page 37
PRESS
the left
arrow key once to change the temperature setpoint of this Step.
Using the numeric keypad enter the number for (from 1 to 71) the temperature of this
Step. Remember that this cabinet has no cooling equipment; therefore, a number entered
into this area that is lower than the actual temperature surrounding the equipment will not
be achieved. This controller works in metric and uses the number input as degrees
Celsius.
PUSH
the enter
key once to save the selection.
Now that the Time and Temperature of this Step are set, the actual test condition must be
set.
PUSH
the NEXT Key
once. The DISPLAY area will show:
PRESS
the left
or the right
arrow key once to position the cursor to change
the Event Code of this Step. Using the numeric keypad enter the number (1, 2, 4, 8,16,
32, 64 or 128) corresponding to the desired Event Code for this Step. As shown
previously (and below), Event Code numbers correspond to actual test conditions.
Description
Event Code
Salt Fog Cycle
1
Event not available on this model
2
Purge 4
Solution Spray (optional)
8
Set current Step to Seconds (not minutes)
16
Guaranteed Soak ON
32
Cover Open / Close (optional)
64
User Output 1
128
PUSH
the enter
key once to save the selection.
Editing
TEST XX(name of test)
Step 1 EVENT XX