BCX Series Operating Manual
XV. TROUBLESHOOTING, NOTES
A. GENERAL
Please note that the controller is a source of information regarding the status of the unit.
Check the basics first
, as with any piece of equipment. Is there electrical power to the
unit or Circuit breaker? Is there water to the unit? Is there compressed air to the unit?
B. SPECIAL NOTE ON SOLUTION PURITY: FILTER, FOG NOZZLE
Attention must be paid to the purity of the chemicals and salt used for various solutions
for tests. A clogged filter and a clogged fog nozzle can be the result of impurities.
Interruption of the fog, uneven collection rates, and improper flow can be caused by these
impurities. Additionally, DI or distilled water that is not of the proper type can invalidate
the test as well as damage the equipment. Never use ASTM D1193 type I or II water.
Use ASTM D1193 Type IV water. Or as described in ISO3696 or JIS K8150.
C. TROUBLESHOOTING CONTENTS
These follow in the order shown below.
No Fog - D.
Low Collection Rates - E.
High Collection Rates - F.
Uneven Collection Rates - G.
Low Relative Humidity - H.
No Heat in Bubble Tower -I.
Cover Bounces - J.
Cloudy Bubble Tower -K.
Incorrect pH Level - L.
Inconsistent Air Pressure in Bubble Tower - M.
Bubble Tower Fills Up with Water Over Float Switch - N.
Salt Settles in Bottom of Salt Solution Reservoir - O