Chapter
9
9.8 Calibration
What does ’
scaling factor
’ mean?
Performing a room calibration results in a certain scaling error. This might be a problem
when applications do have very high accuracy requirements (e.g. when performing mea-
surements with the Measurement Tool ).
The scaling error can be determined, for instance, by measuring points on a defined
scale (e.g. tape measure). The determined scaling error can thus be adjusted with the
DTrack2
’
scaling factor
’ (refer to chapter B.1.1.1 on page 250).
Please contact
ART
if you require further information.
I cannot calibrate my target.
•
Please check if the markers of the target are seen by the cameras in good quality
(i.e. at least yellow) - a slight change of the target position might help. If necessary
increase the flash intensities of the cameras. Therefore, go to
Settings
→
Cameras
.
•
Make sure that when starting the body calibration every marker of the target is visible
to the cameras.
•
A body calibration can be carried out only if a valid room calibration has been carried
out before. Check in the
Event Display
if there is a warning saying ’no valid room
calibration’, additionally the button
Calibrate
is greyed out and cannot be clicked.
In that case, please carry out a new room calibration.
•
Please make sure using the
Monitor 2DOF
display that no marker is overlapped
(i.e. merged) by another one. If necessary rearrange the target and restart the body
calibration.
•
Try to carry out the body calibration with a moving target. Therefore, shift the starting
position of the target for calibration (often a shift of 10 or 20cm is enough to enable
calibration again).
•
Make sure that no other markers or targets are inside the tracking volume other than
the one you want to calibrate.
•
Please carry out a new room calibration.
I’d like to save all my body files for backup reasons but I receive less files than
configured bodies.
DTrack2
only creates body files for calibrated bodies. For example: if you configured 15
targets for tracking but only 4 of them are calibrated, you will only receive 4 calibration
files.
209
Содержание Flystick3
Страница 220: ...10 General Information 10 4 Declaration of conformity 220 ...
Страница 221: ...10 4 Declaration of conformity 221 ...
Страница 222: ...10 General Information 222 ...
Страница 223: ...10 4 Declaration of conformity 223 ...
Страница 224: ...10 General Information 224 ...
Страница 233: ...10 4 Declaration of conformity 233 ...
Страница 234: ...10 General Information 234 ...
Страница 235: ...Appendix 10 4 Declaration of conformity 235 ...
Страница 248: ...A Technical specifications ARTTRACK2 discontinued ARTTRACK3 discontinued TRACKPACK discontinued 248 ...
Страница 249: ...Appendix A A 7 System latency TRACKPACK Cascade discontinued 249 ...
Страница 269: ...Appendix B B 3 Input of Special Control Data via Ethernet tfb 2 0 0 1 0 0 5 0 1 0 8 0 5 269 ...