9 Frequently asked questions (FAQ)
Restart the room calibration. Move the wand gently within the measurement volume, in
order to generate a virtual point cloud. This point cloud should fill at least about two thirds
of the measurement volume.
Please refer to chapter 4.8.3 on page 51 for more information.
I cannot cover two thirds of the measurement volume within the set time for the
room calibration. How do I extend the time for the calibration process?
Please open the dialogue
Calibration
→
Room
and click
Show details
. There you can
extend the duration for the calibration process. Please refer to chapter 8.6.3 on page 166
for more information.
How shall I evaluate the room calibration results?
Please refer to chapter 8.6.3.2 on page 169 for a detailed explanation.
DTrack2
reports the error message "
no angle tool detected
". What went wrong?
First of all, please double-check if all markers are visible and that there are no disturbing
reflexes. Sometimes, it might help to move the calibration angle a little bit.
Please always verify the settings for the room calibration before repeating:
•
wand length is printed on a label on the wand;
•
select the correct calibration set.
The progress bar of the room calibration stopped.
Please double-check in the
Monitor 2DOF
display whether the calibration wand is recog-
nized by the cameras during its movement.
What is a "re-calibration" and when do I have to perform one?
DTrack2
provides simplified room and body calibrations, called room and body re-calibration.
The main advantage of a re-calibration is that
DTrack2
preserves the origin and the ori-
entation of your coordinate system as well as specific information (predefined emission
vectors, cylindrical marker).
You may perform a room re-calibration from time to time or if a camera moved slightly
(e.g. due to mechanical instabilities).Please refer to chapter 8.6.3.3 on page 170.
If your body fell down or if the structure is bent you should carry out a body re-calibration.
Please refer to chapter 8.6.5.7 on page 182 for more information.
208
Содержание Flystick3
Страница 220: ...10 General Information 10 4 Declaration of conformity 220 ...
Страница 221: ...10 4 Declaration of conformity 221 ...
Страница 222: ...10 General Information 222 ...
Страница 223: ...10 4 Declaration of conformity 223 ...
Страница 224: ...10 General Information 224 ...
Страница 233: ...10 4 Declaration of conformity 233 ...
Страница 234: ...10 General Information 234 ...
Страница 235: ...Appendix 10 4 Declaration of conformity 235 ...
Страница 248: ...A Technical specifications ARTTRACK2 discontinued ARTTRACK3 discontinued TRACKPACK discontinued 248 ...
Страница 249: ...Appendix A A 7 System latency TRACKPACK Cascade discontinued 249 ...
Страница 269: ...Appendix B B 3 Input of Special Control Data via Ethernet tfb 2 0 0 1 0 0 5 0 1 0 8 0 5 269 ...