• Bleach, 10% solution
• Tissue, lint-free
• Cotton or nylon swabs and lint-free cloths
• Ethanol, 95% solution
• Safety glasses
• Pipette (100-µL) with pipette tips
• Powder-free gloves
• Screwdriver
• Deionized water
To prevent damaging or contaminating the sample block, handle the assembly as
shown. After the assembly has been removed from the QuantStudio
™
6 or 7 Flex
System, place the sample block on a clean, dry surface or in its shipping container.
WARNING!
PHYSICAL INJURY HAZARD. During instrument operation, the
sample block can reach 100°C. Allow the sample block to cool to room
temperature before performing the procedure.
IMPORTANT!
Wear powder-free gloves when you perform this procedure.
1.
Identify the contaminated wells of the sample block (see “How to identify
2.
Power off and unplug the QuantStudio
™
6 or 7 Flex Real-Time PCR System, then
allow it to cool for 15 minutes.
3.
Open the access door.
Materials required
How to handle the
sample block
Clean the sample
block
Appendix B
Maintenance
Decontaminate the sample block
B
98
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide