4.
Click Next, then prepare for the calibration as instructed.
5.
From the bottom of the Setup tab, enter the reagent information for the ROI plate
or array card that you are using.
6.
Load the calibration plate or array card into the instrument:
a.
From the instrument touchscreen, tap to eject the instrument tray.
b.
Load the plate or array card into the plate holder so that:
• Well A1 of the plate or array card is in the top-left corner of the plate
adapter.
• The barcode faces the front of the instrument.
IMPORTANT!
Plates and array cards should be loaded and unloaded by
operators who have been warned of the moving parts hazard and have been
adequately trained.
c.
From the instrument touchscreen, tap to close the instrument tray.
7.
After loading the plate or array card, start the calibration:
a.
At the bottom of the Setup tab, select Check the box when the ROI
calibration plate has been loaded, then click Next.
b.
In the Run screen, click Start run.
IMPORTANT!
Do not attempt to open the access door during the run. The door
is locked while the instrument is in operation.
Chapter 2
Calibrate the instrument
ROI calibration
2
44
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide