3.
Open the access door.
4.
Remove the lamp from the system:
a.
Slide the lamp release lever downward.
b.
Firmly grasp the lamp and lift it up and out of the slotted mount.
5.
Inspect the lamp for damage (carbon usually coats the inside of a failed lamp). If
necessary, dispose of the lamp and obtain a replacement.
IMPORTANT!
Dispose of the lamp in accordance with your local municipal
waste ordinances. Do not dispose of the lamp as unsorted municipal waste.
6.
Install the new lamp into the QuantStudio
™
6 or 7 Flex Real-Time PCR System:
a.
Slide the lamp release lever upward.
b.
Firmly grasp the lamp, place it into the slotted mount, then carefully slide
the lamp downward into place.
Appendix B
Maintenance
Replace the halogen lamp
B
102
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide