Enable or change the calibration reminders
The calibration reminder settings allow you to configure the QuantStudio
™
Real-Time
PCR Software to alert you by email when the QuantStudio
™
6 or 7 Flex Real-Time
PCR System requires calibration. The notifications settings feature is optional, and it
does not affect instrument performance.
IMPORTANT!
The calibration reminders feature requires a connection between the
QuantStudio
™
6 or 7 Flex Real-Time PCR System computer and a computer network.
If you choose to set up the feature as described below, ensure that your network has
appropriate security measures in place to protect data from threats such as computer
viruses.
IMPORTANT!
The QuantStudio
™
Real-Time PCR Software transmits emails only
while the QuantStudio
™
6 or 7 Flex Real-Time PCR System is monitored. If the
network connection is interrupted, the software will stop transmitting updates.
The QuantStudio
™
Real-Time PCR Software requires access to a Simple Mail Transfer
Protocol (SMTP) server to email calibration reminders. Contact your systems
administrator or information technology department for the following information:
• Network address of an SMTP server.
• A user name and password for the server, if required for access.
• The Secure Sockets Layer (SSL) setting of the server (on or off).
1.
In the QuantStudio
™
Real-Time PCR Software, select Tools
4
Preferences.
2.
From the Preferences dialog box, select the SMTP Settings tab.
3.
Define the settings for the SMTP server:
a.
From the Outgoing Mail Server (SMTP) field, enter the network address of
a Simple Mail Transfer Protocol (SMTP) server.
Optionally, you can specify the transmission control protocol (TCP) port for
the server by appending the port number to the server name, separating the
two using a colon (
:
).
For example:
smtp.mycompany.com:2023
Note:
If a TCP port is not specified, the software uses the default port
number (25).
b.
Select Encryption Required if the mail server has SSL enabled.
c.
Select Authentication Required if the mail server requires a user name and
password.
Requirements to
email calibration
reminders
Define the mail
server settings
Chapter 2
Calibrate the instrument
Enable or change the calibration reminders
2
32
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide