Connect the QuantStudio
™
6 or 7 Flex System to the network
After deciding how to connect the QuantStudio
™
6 or 7 Flex System to a network, set
up the QuantStudio
™
6 or 7 Flex System according to your network policies.
Ethernet cable with RJ45 connectors (a CAT6 Ethernet cable for a 1000Mbit/s network
connection or a CAT5 for 100Mbit/s connection)
1.
Use the Ethernet cable to connect the Ethernet port of the QuantStudio
™
6 or 7
Flex Real-Time PCR System to the nearest network port.
2.
Power on the system.
3.
In the main menu of the touchscreen, tap Settings then tap Set Network
Information.
Note:
The Set Network Information screen displays the Media Access Control
(MAC) address of the system below the Autodiscovery and Smart Monitoring
check boxes. The MAC address can be used to uniquely identify the system on
the network.
4.
Tap Autodiscovery to make the system discoverable by computers that are
running the QuantStudio
™
Real-Time PCR Software.
5.
Tap Smart Monitoring to enable the feature on the QuantStudio
™
6 or 7 Flex
Real-Time PCR System.
The Smart Monitoring feature allows Thermo Fisher Scientific service personnel
to monitor the status of the system remotely through an internet connection.
Smart Monitoring employs multiple layers of security, including a Secure Sockets
Layer (SSL) and Lightweight Directory Access Protocol (LDAP) authentication,
to provide real-time troubleshooting and problem resolution for the instrument.
For a detailed description of the Smart Monitoring Service, see the Smart
Monitoring Service Product Bulletin: Leveraging the power of the Internet while
maintaining system security (Pub. No. 121PB07-03).
6.
Set the Internet Protocol (TCP/IP) Properties for either DHCP or Static IP
communication.
Network service
Action
DHCP
Tap Obtain an IP address automatically.
Static IP
1. Tap Use the following IP address.
2. Tap the IP Address field, enter the IP address using the
keypad, then tap Done.
3. Assign the:
• IP addresses for the DNS Servers (primary and
secondary)
• Subnet Mask setting
• Default Gateway setting
Materials required
Define the
instrument
internet protocol
settings
Appendix C
Networking
Connect the QuantStudio
™
6 or 7 Flex System to the network
C
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide