![Applied Biosystems QuantStudio 6 Flex Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/applied-biosystems/quantstudio-6-flex/quantstudio-6-flex_user-manual_2955112050.webp)
IMPORTANT!
Wear powder-free gloves and safety glasses when you prepare plates
or array cards.
Prepare the calibration consumable appropriate for your QuantStudio
™
6 or 7 Flex
System:
• “Prepare the ROI calibration plate“ on page 50
• “Preparing array cards for instrument calibration“ on page 37
Materials required
Note:
The uniformity calibration requires the same materials used to perform the
ROI calibration.
• Region of Interest (ROI) Plate
• Centrifuge with plate adapter
• Powder-free gloves
• Safety goggles
Prepare the plate
IMPORTANT!
Wear powder-free gloves and safety glasses when you prepare the
plate.
If you have a prepared ROI calibration plate from a recent ROI calibration, go to step
3b; otherwise, prepare the ROI calibration plate as follows:
1.
Remove the ROI calibration plate from the freezer, then thaw it at room
temperature for approximately 30 minutes.
IMPORTANT!
Use the calibration plate within 2 hours of defrosting it. Until you
are ready to run the plate, store it in the dark and at ambient temperature
(15–30°C). Do not remove the plate from its packaging until you are ready to run
it.
2.
Remove the ROI plate from its packaging. Do not remove the optical film.
3.
Vortex and centrifuge the plate:
a.
Vortex the ROI calibration plate for 5 seconds.
b.
Centrifuge the plate for 2 minutes at < 1500 rpm.
IMPORTANT!
The ROI calibration plate must be well mixed and
centrifuged.
Preparing the
calibration plate
or array card
Prepare the ROI
calibration plate
Chapter 2
Calibrate the instrument
Uniformity calibration
2
50
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide