During the ROI calibration, the instrument software captures images of the ROI
calibration plate or array card at each instrument filter. An ROI calibration passes if
the collected image for each filter distinguishes all wells of the ROI plate or array card.
Each well in the image must be distinct and visible at the same luminosity relative to
the other wells in the image.
Note:
You can review the ROI calibration image for each filter set by selecting the
desired filter combination from the Filter Set menu of the ROI tab in the Instrument
Manager.
Passing image
Failing image
Green circles around
all wells indicate that
the wells calibrated successfully. Each
green circle indicates that the region of
interest for the well position is sufficiently
bright.
Red circles around some of the wells
indicate that the wells did not calibrate. The
absence of a circle indicates that the region
of interest for the well position is not
sufficiently bright.
IMPORTANT!
Wear powder-free gloves and safety glasses when you prepare plates
or array cards.
Prepare the calibration consumable appropriate for your QuantStudio
™
6 or 7 Flex
System:
• “Prepare the ROI calibration plate“ on page 42
• “Preparing array cards for instrument calibration“ on page 37
Materials required
• Region of Interest (ROI) Plate
• Centrifuge with plate adapter
• Powder-free gloves
• Safety goggles
About the
calibration data
Preparing the
calibration plate
or array card
Prepare the ROI
calibration plate
Chapter 2
Calibrate the instrument
ROI calibration
2
42
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide