
2-9 Application Interface – Main Menu Tab
Using the 2300-584-R Software with the ME7838A
2-26
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
Default Display Scale Button
This command sets all four channels to a scale of 5 dB/div for the Magnitude displays and to 40 degrees/div for
the Phase displays, and the reference value to 0. This allows you to see if the device is properly connected.
Go To Remote Control Button
This command is only available after you have clicked the
Go to Local Control
button. When you click this
button, the program takes control of the VNA Display type and Display scale settings and sets them to the
program default values.
Go To Local Control Button
This command allows you access to the front panel buttons while the program is running. The instrument is in
Local Lock Out
unless this button is pushed. This prevents an inadvertent front panel button actuation from
affecting the proper operation of the program.
Restart Button
The
Restart
button command allows you to change the tested instrument, the frequency range, or the selected
verification kit. Selecting
Yes
causes the program to return to the program initialization stage.
Usually, this button is used to test for second frequency band and program execution continues in
“Restart for Alternate Frequency Band” on page 2-34
and the
Choose The Restart Method
dialog box appears.
Quit Button
Selecting
Quit
brings up a confirmation dialog box shown in
. Selecting
Yes
exits the program;
selecting
No
returns you to the previous screen.
About
Displays version information, copyright, other legal notices, and company contact information.
Run All Tests
This command button runs both the Matched Thru and Mismatched Thru tests. Program execution starts with
Section 2-11 “Matched Thru Tests” on page 2-29
. When both tests are completed, program execution returns to the
Main Menu tab
shown in
Figure 2-36, “ME7838A Application Interface – Main Menu Tab” on page 2-24
.
Caution
Modifying the VNA settings (such as power level, bandwidth, averaging, etc.) could adversely affect
the measurement results. This should only be used to view the data or for using markers.
Figure 2-38.
Confirmation Dialog
Содержание 3656B W1
Страница 2: ......
Страница 12: ...Contents 10 PN 10410 00286 Rev L 3656B W1 Cal Ver Kit 2300 584 R PVS UG...
Страница 277: ......