Using the 2300-584-R Software with the ME7838E
3-8 High-Band Calibration
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
3-21
2.
When ready to proceed, click
OK
. The
VNA Calibration: Step 1 of 5
dialog box appears.
a.
Connect the Male Cal Kit Isolation Device to VNA Port 1.
b.
Connect the Female Cal Kit Isolation Device to VNA Port 2.
c.
When ready to proceed, click
OK
.
3.
The
VNA Calibration: Step 2 of 5
dialog box appears.
a.
Connect the Male 2.02 mm Short #1 to VNA Port 1.
b.
Connect the Female 2.65 mm Short #2 to VNA Port 2.
c.
When ready to proceed, click
OK
.
4.
The
VNA Calibration: Step 3 of 5
dialog box appears.
a.
Connect the Male 2.65 mm Short #2 to VNA Port 1.
b.
Connect the Female 3.18 mm Short #3 to VNA Port 2.
c.
When ready to proceed, click
OK
.
Figure 3-30.
VNA Calibration: Step 1 of 5 Dialog
Figure 3-31.
VNA Calibration: Step 2 of 5 Dialog
Figure 3-32.
VNA Calibration: Step 3 of 5 Dialog