3-9 Application Interface – Main Menu Tab
Using the 2300-584-R Software with the ME7838E
3-26
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
Repeat Last Cal
This command button re-calibrates the VNA according to the CURRENT calibration. This provides a way to
recalibrate the VNA without having to reset the program parameters in case the original calibration was
flawed for some reason. When the calibration sequence is complete, program execution continues to the
Main
Menu tab
shown previously in
Figure 3-35, “ME7838E Application Interface – Main Menu Tab” on page 3-23
.
Save Data
This command button writes data files to the PC Controller and stores the files at
C:\mmdcsvc\Data
. The
program does not write data to the verification kit USB memory device. If selected, program execution
continues as described in
Section 3-13 “Saving Verification Data” on page 3-33
The number of saved files varies depending on the user settings:
•
If both the matched and mismatched thru tests are run, two files are written to the PC Controller hard
drive.
•
If only one of the tests was run, only one file is written to the PC Controller’s hard drive at
C:\mmcsvc\Data
Matched Thru
This command button only runs the Matched Thru tests. If selected, program execution continues as described
in
Section 3-11 “Matched Thru Tests” on page 3-28
. When the tests are completed, the program returns to the
Main Menu tab
shown in
Section 3-9 “Application Interface – Main Menu Tab” on page 3-23
Matched Thru SN
This field is used to enter the serial number of the matched thru.
Mismatched Thru
This command button only runs the Mismatched Thru tests. If selected, program execution continues in
Section 3-12 “Mismatched Thru Tests” on page 3-30
. When the tests are completed, the program returns to the
Main Menu tab
shown in
Figure 3-35, “ME7838E Application Interface – Main Menu Tab” on page 3-23
.
Mismatched Thru SN
This field is used to enter the serial number of the matched thru.