5-5 Running the Verification Software
Using the 2300-584-R Software with the ME7838E4
5-6
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
5-5
Running the Verification Software
After the VNA and PC have been configured as described above, you are ready to run the program and perform
the calibration/verification. Make sure you have the following information available:
•
Serial number of the 3656B W1 Calibration/Verification Kit
•
Serial number of the 3739C Broadband/Millimeter Wave Test Set
•
Serial Number of the 3736B Master Test Set
•
MN469xC 4-port Test Set
•
Model number and serial number of the Port 1 and Port 2 BB/mm Module
•
Model number and serial number of the Port 3 and Port 4 BB/mm Module
Procedure
1.
With the equipment and software configured as described in
“Configuring the Hardware” on page 5-1
turn on the computer and allow it to start Windows.
2.
Locate the
Anritsu W1 Verification
icon (shown below) on the desktop and double-click it.
3.
The program displays an
About
dialog box with version information. Click
OK
to continue.
4.
The
Test Operator Name
dialog box appears. Enter a user name or other identification.
Figure 5-6.
About Dialog
Figure 5-7.
Name Dialog