Using the 2300-584-R Software with the ME7838E4
5-8 High-Band Calibration Port 1 – Port 2
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
5-25
3.
The
VNA Calibration: Step 2 of 5
dialog box appears.
a.
Connect the Male 2.02 mm Short #1 to VNA Port 1.
b.
Connect the Female 2.65 mm Short #2 to VNA Port 2.
c.
When ready to proceed, click
OK
.
4.
The
VNA Calibration: Step 3 of 5
dialog box appears.
a.
Connect the Male 2.65 mm Short #2 to VNA Port 1.
b.
Connect the Female 3.18 mm Short #3 to VNA Port 2.
c.
When ready to proceed, click
OK
.
Figure 5-39.
VNA Calibration: Step 2 of 5 Dialog
Figure 5-40.
VNA Calibration: Step 3 of 5 Dialog