4-22 Restart for Alternate Frequency Band
Using the 2300-584-R Software with the ME7838A4
4-60
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
4-22 Restart for Alternate Frequency Band
After the
Matched Thru
and
Mismatched Thru
tests have been successfully completed and the data saved, it is
necessary to restart the program and test the other frequency band. For example, if you tested for
40 MHz to
65 GHz Frequency Range
on your first calibration pass, on the second pass, select the
65 GHz to 110 GHz
Frequency Range
.
Procedure
1.
Ensure that the verification results of the previous test have been saved to the
C:\mmdscvc\Data
.
2.
To perform the verification tests for the second frequency range, on the
System Verification Software Main
Menu - Main Menu Tab,
select the
Restart
button. The main menu is described in
Section 4-17, “Application Interface – Main Menu Tab” on page 4-47
.
3.
The
Choose The Restart Method
dialog box appears as shown in
4.
If you selected
Full Restart
, the calibration process starts again and the procedure steps are the same as
described previously starting in
Section 4-5, “Running the Verification Software” on page 4-6
,
.
5.
On the
Setup Menu
, repeat the data entry for the following items. Click
OK
to proceed to the next step.
6.
Enter the
GPIB Board 0.
7.
Enter the
GPIB Address 6
.
8.
Enter the
Test Set Serial Number.
9.
Enter the
User Name
.
10.
Enter the
Cal Kit Type
selection.
11.
Enter the
Cal Kit Serial Number
selection.
12.
Enter the
Load Cal Kit Disk from VNA Floppy
selection.
Figure 4-83.
Choose The Restart Method Dialog