Using the 2300-584-R Software with the ME7838A
2-6 Application Interface – Setup Menu Tab
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
2-11
2-6
Application Interface – Setup Menu Tab
The
SPV Application Interface
dialog box displays three tabs with configuration, control, and serial number
information. Each tabbed dialog box contains buttons to control program operations as outlined in the
following sections:
•
The
Setup Menu
tab is described in this section.
•
The
Main Menu
tab display is described in
Section 2-9, “Application Interface – Main Menu Tab”
.
•
The
Serial Number Info
tab is described in
Section 2-10, “Application Interface – Serial Number Tabs”
.
Procedure
The following procedure continues the program setup from the previous section:
1.
In the
Calibration Kit Type
area, select
W Coax
.
2.
In the
Serial Number (5 Digits)
area, enter the serial number for the 3656B Calibration/Verification Kit. At
least five (5) digits are required.
3.
In the
Load Cal-Kit Coefficients to VNA
area, select
Yes
.
4.
The
Install Calibration-Kit Coefficients to the VNA
dialog box appears.
Figure 2-15.
ME7838A Application Interface – Setup Menu Tab
W Coax
Cal-Kit