Maintenance
9-2 Visual Inspection
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
9-3
Center Conductor Inspection
Inspect for bent, broken, or damaged center conductors:
Inspect for center pin concentricity:
Index
Description
1
New condition
2
Normal use – Slight gap
3
Bent pin section – Discard connector
4
Fingers spread or collapsed – Discard connector
5
Broken pin or pin section – Discard connector
Figure 9-2.
Example Connector Damage States
Index
Description
1
Centered
2
Maximum allowable off-center
3
Fail – off center – Discard connector
Figure 9-3.
Example Connector Concentricity
3
Bent
5
Broken