4-6 Application Interface – Setup Menu Tab
Using the 2300-584-R Software with the ME7838A4
4-16
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
7.
The
Application Interface
dialog box reappears displaying the
Setup Menu
tab and
Frequency Range
area.
In the
Frequency Range
area, select either the low range (40 MHz to 65 GHz) or the high range (65 GHz
to 110 GHz). For purposes of this document, the low band is done first. The range that is not selected is
calibrated during the Restart Procedure described in
Section 4-14 “Restart for Alternate Frequency
. After selecting the frequency range, click OK. The
Verification Kit Information
area
appears.
Figure 4-26.
ME7838A4 Application Interface – Setup Menu Tab