2300-584-R Software with the Lightning ME7808A/B/C
7-10 Troubleshooting
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
PN: 10410-00286 Rev. L
7-15
i.
Enter the
Frequency Range
selection. Select the frequency range that has not been tested. If the
first calibration procedure selected the 40 MHz to 65 GHz frequency range, at this point, select the
65 GHz to 110 GHz frequency range.
j.
Enter the
Verification Kit Type
selection.
k.
Enter the
Verification Kit Serial Number
selection.
l.
Enter the
PC Floppy Drive the Disk Is In
selection.
46.
Complete the calibration and verification of the devices for the second frequency range.
47.
Save the second set of calibration data and verification results to the floppy disk.
48.
This series of automated tests verify that the ME7808A/B/C broadband S-parameter measurement
accuracy meets factory specifications.
49.
This ends the calibration and verification procedure for the Lightning ME7808A/B/C.
7-10 Troubleshooting
Difficulty Running the Program
If you have difficulty getting the program to run properly:
1.
Check your GPIB interconnection cables and addresses.
2.
Check to see that the Windows GPIB is present on the boot drive, that it is properly configured, and that
it passes the National Instruments hardware and software tests.
3.
This version of the verification software must be installed with the install program on the USB memory
device. The program does not run if it is just copied from the USB memory device to the hard disk.
4.
Ensure that, after starting-up the performance verification software, the verification kit data disk is
installed in drive A or B and that it contains 20 files.
If, after checking the above, you are still having difficulty, contact your Anritsu customer service center and
ask for the Vector Network Analyzer support engineer for further assistance.
Difficulty Meeting System Specifications
If the verification software appears to run properly, but the results are not within the measurement limits
associated with the verification kit:
1.
Check both the verification kit and calibration kit devices for signs of physical damage. Make sure that
the connectors are clean.
2.
Ensure that the serial number of the verification kit data disk matches that shown on the verification kit.
3.
Repeat the process with a fresh calibration. Save the results of both measurements as an aid in
troubleshooting, in case you require factory assistance.
4.
When installing calibration devices, and when measuring verification devices, pay particular attention to
proper connector alignment and torque. Torque the connector using the torque wrench supplied with the
calibration kit.
5.
Assure all active systems have been powered on at least one hour before the start of the calibrations.
If you still have difficulty after following the above steps, please contact Anritsu Customer Service and ask for
the Vector Network Analyzer support engineer for further assistance.