3-14 Restart for Alternate Frequency Band
Using the 2300-584-R Software with the ME7838E
3-34
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
Procedure
1.
Ensure that the verification results of the previous test have been saved to the
C:\mmdscvc\Data
.
2.
To perform the verification tests for the second frequency range, on the
System Verification Software Main
Menu - Main Menu Tab,
select the
Restart
button. The main menu is described in
Interface – Main Menu Tab” on page 3-23
.
3.
The
Choose The Restart Method
dialog box appears as shown in
4.
If you selected
Full Restart
, the calibration process starts again and the procedure steps are the same as
the procedures described previously starting in
Section 3-5, “Running the Verification Software”
.
5.
On the
Setup Menu
, repeat the data entry for the following items. Click
OK
when ready to proceed to the
next step.
6.
Enter the
GPIB Board 0.
7.
Enter the
GPIB Address 6
.
8.
Enter the
Test Set Serial Number.
9.
Enter the
User Name
.
10.
Enter the
Cal Kit Type
selection.
11.
Enter the
Cal Kit Serial Number
selection.
12.
Enter the
Load Cal Kit Disk from VNA Floppy
selection.
13.
Enter the
Frequency Range
selection. Select the frequency range that has not been tested.
a.
If the first calibration procedure selected the
40 MHz to 65 GHz Frequency Range
, select the
65 GHz to 110 GHz Frequency Range
.
b.
If the first calibration procedure selected the
65 GHz to 110 GHz Frequency Range
, select the
40 MHz to 65 GHz Frequency Range
.
Figure 3-49.
Choose The Restart Method Dialog