7-8 Testing the Verification Standards
2300-584-R Software with the Lightning ME7808A/B/C
7-12
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
7-8
Testing the Verification Standards
Whenever the program is executing a command from the main menu, the buttons are disabled until the
currently executing command is finished. This is necessary due to the event driven nature of Windows. Also,
while a command is being executed, informational messages may be displayed on the screen.
30.
Click either the
Matched Thru
or
Mismatched Thru
command buttons and follow the on-screen
instructions.
31.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
Figure 7-16.
Anritsu 37397 Verification – Main Menu Tab
Figure 7-17.
Connect the Matched Thus
Note
The verification standards must be connected to the W1 coupler ports and not to the VNA ports.