2-13 Saving Verification Data
Using the 2300-584-R Software with the ME7838A
2-34
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
2-13 Saving Verification Data
If the
Save Data
button on the
Main Menu
tab was selected, the save data dialogs described below appear.
When the save data procedure is completed, the program returns to the main menu as described previously in
Section 2-9 “Application Interface – Main Menu Tab” on page 2-24
.
The default directory path and file names are:
•
C:\mmdcsvc\Data\Matched Thru Low.dat
•
C:\mmdcsvc\Data\Mismatched Thru Low.dat
•
C:\mmdcsvc\Data\Matched Thru High.dat
•
C:\mmdcsvc\Data\Mismatched Thru High.dat
If the default file names should be changed, select
Yes
and enter the new file names. Otherwise, select
No
.
The following dialog is shown when the files are successfully saved to disk.
2-14 Restart for Alternate Frequency Band
After the
Matched Thru
and
Mismatched Thru
tests have been successfully completed and the data saved, it is
necessary to restart the program and test the other frequency band. For example, if you tested for
40 MHz to
65 GHz Frequency Range
on your first calibration pass, on the second pass, select the
65 GHz to 110 GHz
Frequency Range
.
Figure 2-48.
Confirmation Dialog
Figure 2-49.
Information Dialog