Table 6-14. Standard Event Status Register Definitions
Bit
Weight
Description
PON (7)
128
Power on bit
1 is set to this bit when the power is turned on.
When this register is read, this bit is cleared to 0, and it remains 0 until the
power is turned on again.
URQ (6)
64
User request bit
Always 0 (not used)
CME (5)
32
Command error
1 is set to this bit when a syntax error occurs in the program code.
EXE (4)
16
Execution error
1 is set to this bit when a parameter is out of the setting range or when
settings conflict.
DDE (3)
8
Device definition error
Always 0 (not used)
QYE (2)
4
Query error
1 is set to this bit when there is no data in the buffer containing response
messages when trying to read that buffer, or when the data in the buffer
containing response messages has been lost.
RQC (1)
2
Request control
Always 0 (not used)
1 is set to this bit when all processing of operation complete (OPC)
commands has been completed.
In this device, this bit is always 0.
OPC (0)
1
Operation complete
1 is set to this bit when all processing of operation complete (OPC)
commands has been completed.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......