User Manual
EC1000S
California Instruments
242
6.6 Error Message List
USB error messages are listed in
Table 6-15. USB Error Message List
No.
Message
Cause
0
“No error”
No error.
100
“Command error”
Command was recognized as invalid due to any cause.
(General command error)
101
“Invalid character”
Command was recognized as invalid due to invalid
characters (such as “VOLT&”) in command string.
102
“Syntax error”
Command was recognized as invalid due to reception of
undefined command or parameter.
200
“Execution error”
Error occurred when executing the command due to any
cause. (General execution error)
222
“Data out of range”
Error occurred when executing the command due to out of
range parameter.
350
“Too Many Errors”
New errors cannot be retained due to error queue overflow.
1
“Invalid with output on”
Error occurred when executing the command because the
output relay was on.
2
“Invalid with output off”
Error occurred when executing the command because the
output relay was off.
3
“Invalid in this mode”
Error occurred when executing the command due to
inappropriate output mode.
4
“Under Busy State”
Error occurred during execution of command due to busy
device.
5
“Out of Limiter”
Error occurred during execution of command because
limiter value was exceeded.
6
“Invalid in this frequency”
Error occurred when executing the command due to
inappropriate oscillation frequency setting.
7
“Out of Input Signal
frequency”
Error occurred when executing the command due to out of
the external synchronization frequency range.
8
“Under Warning State”
Error occurred when executing the command because
device was operating in protected mode.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......