User Manual
EC1000S
California Instruments
80
4.1 Use as a DC Power Source
When using the EC1000S as a DC voltage source, select DC - Internal signal mode (AC+DC-INT), then
enter the settings listed in
AC components (sine wave, square wave, and arbitrary waveform) can be superimposed on the DC
power source.
A protection diode may be required to be connected, depending on the load, such as a capacitor or
inductor.
Table 4-1. Panel Settings When Using EC1000S as a DC Power Source
Setting
Setting Range
Resolution
Initial Value
DC voltage
100
V
range
200.0 to +200.0 V
0.1 V
0.0 V
200
V
range
400.0 to +400.0 V
0.1 V
0.0 V
AC voltage
100
V
range
SIN/SQU
0.0 to 140.0 Vrms
0.1 Vrms
0.0 Vrms
ARB1 to ARB16
0.0 to 400.0 Vp-p
0.1 Vp-p
0.0 Vp-p
200
V
range
SIN/SQU
0.0 to 280.0 Vrms
0.1 Vrms
0.0 Vrms
ARB1 to ARB16
0.0 to 800.0 Vp-p
0.1 Vp-p
0.0 Vp-p
AC voltage frequency
1.0 to 550.0 Hz
0.1 Hz
50.0 Hz
Phase when output is on
0.0 to 359.9
0.1
0.0
The unit for AC voltage settings differs according to the AC voltage waveform setting.
4.1.1 Connecting output terminal to load during DC output
Use the output terminal on the rear panel when outputting DC+AC or only DC from the EC1000S.
If the load is a capacitor or inductive load, connect a diode for protection that meets the following
specifications between the output terminal of the EC1000S and the load.
Maximum reverse voltage: 600 V or higher
Maximum forward current: 15 A or more (100 V range), 7.5 A or more (200 V range)
a) Capacitive load such as capacitor (about 50
F or more) and DC-DC converter
When a capacitive load is charged, the greater the voltage changes, the more the charge
current becomes. Overshoot may be generated in the output currents depending on the load,
even if the peak current limiter of the EC1000S operates, thus turning off the output due to
overcurrent protection.
To avoid this, gradually lower the set voltage so that decrease of the voltage changes per certain
unit time or use the sweep function of the sequence.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......