User Manual
EC1000S
California Instruments
163
See
, for description of waveform settings.
S
ee the control software user’s manual
, for description of arbitrary waveform transfers.
5.3.7 Output peak current limiter settings
The positive value and the negative value of the peak current limiter can be set independently.
If the output peak current exceeds the set current, the output voltage is clipped.
Setting range limits can be set for output voltage and output frequency.
See “5.5.5 Setting range limiter (LIMIT)”
.
Table 5-16. Output Peak Current Limiter Setting Range
Setting
Setting Range
Resolution Initial Value
Output peak current
limiter (positive)
100 V range
+10.0 A to +42.0 A
0.1 A
+42.0 A
200 V range
+5.0 A to +21.0 A
0.1 A
+21.0 A
Output peak current
limiter (negative)
100 V range
42.0 A to
10.0 A
0.1 A
42.0 A
200 V range
21.0 A to
5.0 A
0.1 A
21.0 A
Limiter settings are retained separately for the 100 V and 200 V range.
Since these are not fully constant current operations, the specified limit value may not be achieved,
depending on the load.
Operation steps
Select the icon in the SET menu, then set a numerical value.
Output peak current limiter (positive):
Output peak current limiter (negative):
S
ee “3.4.7 Using limiter functions”
, for description of the output peak current limiter settings.
Note with caution that the output peak current limiter may allow overshooting, depending on the load,
and therefore may not achieve the peak current value as set.
For an inductive load, operation of the limiter can cause an output overvoltage, triggering the
protection function. In this case, the peak current limiter settings (positive and negative) should be
reduced.
!
CAUTION
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......