4.2 Measurement of Inrush Current
4.2.1 Inrush current
In such an electrical product as contains a capacitor input type rectifier circuit for the input power line,
a current that is larger than the rated current may flow for a short time immediately after the power
source switch is turned on.
This current is called an “inrush current.”
If the power source is not able to supply such a large current, enough power is not supplied to the
product, and sometimes it does not start in the worst case.
On the other hand, when such a large current passes through the power source line, the supply
voltage is likely to decrease because of the voltage drop at the line impedance. There exists a
standard to limit this effect within a certain level.
Such a product should be driven by a power source with enough supply capacity
Example of an inrush current
An electric drill
(rated 100 V AC/380 W)
Upper: Voltage (100 V/div)
Lower: Inrush current (10 A/div)
Even in a small electric drill, an inrush current of 15 Apk flows right after the power source is turned
on. Since the rated current is 3.8 A, this represents a fourfold increase in the current flow.
The EC1000S can supply four times as large peaked current as the rated current supporting such a
usage. There is no need to use a power source with an extra large capacity in order to supply the
inrush current (for a capacitor input circuit of CF = 4 or less).
The initial phase of the output AC voltage is adjustable from 0.0
to 359.9
(in 0.1
steps), so the
user can monitor how the current rushes in the load under various phase conditions. The peak
current hold measurement function enables the measurement of the maximum peak value for the
current flowing at that time.
When measuring the inrush current, set the peak current limiter to the maximum value. If the limiter
setting is lowered, the inrush current measurement may be incorrect.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......