User Manual
EC1000S
California Instruments
251
7.2.2 Protection function-related errors
When a protection function-related error has occurred, an error message is displayed in the message
window, and operation is stopped.
Figure 7-2. Screen Display When Protection Function-Related Error Has Occurred
When a protection function-related error message is displayed, turn off the power switch. (
See
.)
lists error messages and actions when errors are detected.
Table 7-3. Protection Function-Related Errors
Error Message
Protection Item Cause and Operation or Action Required
When Fault Is Detected
System Locked.
caused by:
DCPS FAILURE
Internal power
source error
Output is set to off and system lock status
is set. Turn off the power switch.
If the same protection error occurs again
after the power is turned back on, contact
California Instruments or its
representative.
System Locked.
caused by:
COMMUNICATION FAILURE 1
Internal
communication
error 1
System Locked.
caused by:
COMMUNICATION FAILURE 2
Internal
communication
error 2
System Locked.
caused by:
FRONT PANEL FAILURE
Control panel
error
System Locked.
caused by:
UNKNOWN ERROR
Unexpected
error
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...
Страница 248: ......
Страница 266: ......
Страница 274: ......