User Manual
EC1000S
California Instruments
100
4.4 Using the Sequence Function
4.4.1 Sequence operations
When the sequence function of the EC1000S is used, the pattern to be output can be freely
programmed.
For example, if a series of voltage change patterns that will be used in voltage variation tests for car
electronics are programmed in advance, the output will change according to the sequence, so the test
can be completed in one series of operations.
In the sequence operation programming, DC, sine waves, and square waves, which are built in the
EC1000S, and arbitrary waveform (16 types) captured via the USB interface are able to use. Maximum
step number is 255. In each step, waveform, level, and duration time can be set and
constant/keep/sweep can be chosen. In addition, sequences can be repeated from 1 to 999 times (or
continuously). The EC1000S is also equipped with other functions, such as for branching to a specified
step during a sequence operation, and for starting, stopping, or holding a sequence operation.
Data is saved to sequence memory automatically.
These sequence functions can be used in AC-INT mode or AC+DC-INT mode.
No. of sequences:
Retained for each operation mode (AC/AC+DC) and each output voltage
range (100 V/200 V).
No. of steps:
1 to 255 (per sequence)
Step time:
0.1 ms to 999.9999 s (resolution 0.1 ms or 0.0001 s)
Operations within
step:
Constant, keep, or linear sweep
Parameter:
DC voltage
Note
, AC voltage
Note
, frequency, waveform, phase (start, end),
step synchronized output (2 bits)
Jump times:
1 to 999 or continuous
Sequence control:
Start:
Starts a sequence.
Stop:
Stops a sequence.
Hold:
Maintains current settings. This is restarted when
a sequence is started.
Branch:
Branches to the specified step.
Note
The setting ranges for AC voltage and DC voltage are determined by the output voltage
ranges set when editing the sequence data.
Sequence control cannot be executed when output is off.
Содержание EC1000S
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 10: ...User Manual EC1000S 6 9 SPECIFICATIONS Lists the specifications functions and performance of the EC1000S...
Страница 18: ...User Manual EC1000S Contents California Instruments 14 9 17 External Dimensions and Weight 286...
Страница 24: ......
Страница 32: ...User Manual EC1000S California Instruments 28...
Страница 41: ...User Manual EC1000S California Instruments 37 The product itself has a withstand voltage of 1500 V AC...
Страница 46: ......
Страница 145: ...User Manual EC1000S California Instruments 141 ENTER CANCEL...
Страница 146: ......
Страница 194: ......
Страница 217: ...User Manual EC1000S California Instruments 213 Response example 200 0...