Math Waveforms
4
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
99
1
Press the
Display Math
softkey and turn the Entry knob to select the
math function you want to display. Then, either push the Entry knob or
press the
Display Math
softkey again to display the selected math
function.
2
Press the
[Math]
key. Then, press the
Display Math
softkey and select the
math function you want to use. Then, press the
Operator
softkey and
select
FFT
.
•
Source 1
— selects the source for the FFT.
•
Span
— sets the overall width of the FFT spectrum that you see on
the display (left to right). Divide span by 10 to calculate the number
of Hertz per division. It is possible to set Span above the maximum
available frequency, in which case the displayed spectrum will not
take up the whole screen. Press the
Span
softkey, then turn the Entry
knob to set the desired frequency span of the display.
•
Center
— sets the FFT spectrum frequency represented at the center
vertical grid line of the display. It is possible to set the Center to
values below half the span or above the maximum available
frequency, in which case the displayed spectrum will not take up the
whole screen. Press the
Center
softkey, then turn the Entry knob to
set the desired center frequency of the display.
•
Scale
— lets you set your own vertical scale factors for FFT expressed
in dB/div (decibels/division). See
"To adjust the math waveform scale
and offset"
on page 91.
•
Offset
— lets you set your own offset for the FFT. The offset value is
in dB and is represented by the center horizontal grid line of the
display. See
"To adjust the math waveform scale and offset"
on
page 91.
•
More FFT
— displays the More FFT Settings Menu.
3
Press the
More FFT
softkey to display additional FFT settings.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...