Waveform Generator
17
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
307
To restore waveform generator defaults
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
3
In the Waveform Generator Settings Menu, press the
Default Wave Gen
softkey.
The waveform generator factory default settings (1 kHz sine wave,
500 mVpp, 0 V offset, High- Z output load) are restored.
To set up dual channel tracking
You can set up one waveform generator output to track adjustments in the
other waveform generator output.
To set up dual channel tracking:
1
Press the
[Wave Gen1]
or
[Wave Gen2]
key for the waveform generator
output that you want to be tracked.
2
In the Waveform Generator Menu, press
Settings
.
3
In the Waveform Generator Settings Menu, press
Dual Chan
.
4
In the Waveform Generator: Dual- Channel Menu, you have these
options:
•
Tracking
— frequency, amplitude, offset, and duty cycle adjustments to
this waveform generator output signal are tracked by the other
waveform generator output.
•
Freq Tracking
— frequency adjustments to this waveform generator
output signal are tracked by the other waveform generator output.
•
Amp Tracking
— amplitude and offset adjustments to this waveform
generator output signal are tracked by the other waveform generator
output.
•
Copy Waveform to WaveGen2/1
— sets the other waveform generator
output to be identical to this waveform generator output (except that
the shape of either output may be inverted).
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...