![Agilent Technologies DSO-X 4022A Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/agilent-technologies/dso-x-4022a/dso-x-4022a_user-manual_2868638076.webp)
76
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
2
Horizontal Controls
•
Captured data (see
"To navigate time"
on page 76).
•
Search events (see
"To navigate search events"
on page 76).
•
Segments, when segmented memory acquisitions are turned on (see
"To
navigate segments"
on page 77).
You can also access navigation controls on the touchscreen. See
"Select
Sidebar Information or Controls"
on page 47.
To navigate time
When acquisitions are stopped, you can use the navigation controls to play
through the captured data.
1
Press
[Navigate]
.
2
In the Navigate Menu, press
Navigate
; then, select
Time
.
3
Press the
navigation keys to play backward, stop, or play
forward in time. You can press the
or
keys multiple times to
speed up the playback. There are three speed levels.
You can also access navigation controls on the touchscreen. See
"Select
Sidebar Information or Controls"
on page 47.
To navigate search events
When acquisitions are stopped, you can use the navigation controls to go
to found search events (set using the
[Search]
key and menu, see
"Searching for Events"
on page 74).
1
Press
[Navigate]
.
2
In the Navigate Menu, press
Navigate
; then, select
Search
.
3
Press the
back and forward keys to go to the previous or next
search event.
When searching Serial decode:
•
You can press the
stop key to set or clear a mark.
•
The
Auto zoom
softkey specifies whether the waveform display is
automatically zoomed to fit the marked row as you navigate.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...