Mask Testing
Mask Test Setup Options
From the Mask Test Menu, press the
Setup
softkey to enter the Mask Setup
Menu.
Run Until
The Run Until softkey lets you specify a condition on which to terminate
testing.
• Forever — The oscilloscope runs continuously. However, if an error occurs
the action specified using the On Error softkey will occur.
• Minimum # of Tests — Choose this option and then use the # of Tests
softkey to select the number of times the oscilloscope will trigger, display
the waveform(s), and compare them to the mask. The oscilloscope will
stop after the specified number of tests have been completed. The
specified minimum number of tests may be exceeded. If an error occurs the
action specified using the On Error softkey will occur. The actual number of
tests completed is displayed above the softkeys.
• Minimum Time — Choose this option and then use the Test Time softkey
to select how long the oscilloscope will run. When the selected time has
passed the oscilloscope will stop. The specified time may be exceeded. If
an error occurs the action specified using the On Error softkey will occur.
The actual test time is displayed above the softkeys.
• Minimum Sigma — Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the
oscilloscope will perform the action specified by the On Error softkey.)
Note that this is a test sigma (the max achievable process sigma,
assuming no defects, for a certain number of tested waveforms) as
opposed to a process sigma (which is tied to the amount of failures per
test). The sigma value may exceed the selected value when a small sigma
value is chosen. The actual sigma is displayed.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...