Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
3
Vertical Controls
To specify the probe attenuation
This is set automatically if the oscilloscope can identify the connected
probe. See Analog channel inputs (see
page 42
).
The probe attenuation factor must be set properly for accurate
measurement results.
If you connect a probe that is not automatically identified by the
oscilloscope, you can manually set the attenuation factor as follows:
1
Press the channel key.
2
Press the
Probe
softkey until you have selected how you want to specify
the attenuation factor, choosing either
Ratio
or
Decibels
.
3
Turn the Entry knob
to set the attenuation factor for the
connected probe.
When measuring voltage values, the attenuation factor can be set from
0.1:1 to 10000:1 in a 1- 2- 5 sequence.
When measuring current values with a current probe, the attenuation
factor can be set from 10 V/A to 0.0001 V/A.
When specifying the attenuation factor in decibels, you can select values
from - 20 dB to 80 dB.
If Amps is chosen as the units and a manual attenuation factor is chosen,
then the units as well as the attenuation factor are displayed above the
Probe
softkey.
To specify the probe skew
When measuring time intervals in the nanoseconds (ns) range, small
differences in cable length can affect the measurement. Use
Skew
to
remove cable- delay errors between any two channels.
1
Probe the same point with both probes.
2
Press one of the probes associated channel key.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...