Waveform Generator
17
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
299
All points are adjusted to maintain uniform time spacing between
points.
Capturing Other Waveforms to the Arbitrary Waveform
The Edit Waveform Menu is opened by pressing
Edit Waveform
in the main
Waveform Generator Menu.
To capture another waveform to the arbitrary waveform:
1
Press
Source
; then, use the Entry knob to select the analog channel,
math, or reference location whose waveform you wish to capture.
2
Press
Store Source to Arb
.
The source waveform is decimated into 8192 (maximum) or fewer
arbitrary waveform points.
To output the waveform generator sync pulse
1
If the Waveform Generator Menu is not currently displayed on the
oscilloscope's softkeys, press the
[Wave Gen]
key.
2
In the Waveform Generator Menu, press the
Settings
softkey.
C A U T I O N
When you create a new arbitrary waveform, the existing arbitrary waveform definition
is overwritten. Note that you can use the [Save/Recall] key and menu to save
arbitrary waveforms to one of four internal storage locations or to a USB storage
device, and you can recall them later. See
"To save arbitrary waveforms"
on page 316
and
"To recall arbitrary waveforms"
on page 320.
N O T E
If the source waveform frequency and/or voltage exceed the capabilities of the waveform
generator, the arbitrary waveform will be limited to the capabilities of the waveform
generator. For example, a 20 MHz waveform captured as the arbitrary waveform, becomes a
12 MHz waveform.
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...