I2C/SPI Triggering and Serial Decode
25
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
417
3
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob
to select the serial slot (Serial 1 or Serial 2) on which the I
2
C signals
are being decoded.
4
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
Start Condition
— The oscilloscope triggers when SDA data transitions
from high to low while the SCL clock is high. For triggering purposes
(including frame triggers), a restart is treated as a start condition.
•
Stop Condition
— The oscilloscope triggers when data (SDA) transitions
from low to high while the clock (SCL) is high.
•
Missing Acknowledge
— The oscilloscope triggers when SDA data is high
during any Ack SCL clock bit.
•
Address with no Ack
— The oscilloscope triggers when the acknowledge
of the selected address field is false. The R/W bit is ignored.
•
Restart
— The oscilloscope triggers when another start condition occurs
before a stop condition.
•
EEPROM Data Read
— The trigger looks for EEPROM control byte value
1010xxx on the SDA line, followed by a Read bit and an Ack bit. It
then looks for the data value and qualifier set by the
Data
softkey
and the
Data is
softkey. When this event occurs, the oscilloscope will
trigger on the clock edge for the Ack bit after the data byte. This
data byte does not need to occur directly after the control byte.
SDA
SCL
Stop
Condition
R/
Start
Condition
Ack
Ack
Data
Address
Содержание DSO-X 4022A
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 8: ...8 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 26: ...26 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide...
Страница 62: ...62 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 1 Getting Started...
Страница 78: ...78 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 118: ...118 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 4 Math Waveforms...
Страница 142: ...142 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Страница 148: ...148 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Страница 156: ...156 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 8 Display Settings...
Страница 204: ...204 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 10 Triggers...
Страница 240: ...240 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 13 Cursors...
Страница 270: ...270 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 14 Measurements...
Страница 286: ...286 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Страница 322: ...322 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 18 Save Recall Setups Screens Data...
Страница 328: ...328 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 19 Print Screens...
Страница 362: ...362 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 366: ...366 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 21 Web Interface...
Страница 404: ...404 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 23 CAN LIN Triggering and Serial Decode...
Страница 414: ...414 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 24 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Agilent InfiniiVision 4000 X Series Oscilloscopes User s Guide 29 USB 2 0 Triggering and Serial Decode...